38 LogTag Analyzer User Guide (2.0)
Function
Purpose
Readings above X
Readings below Y
(inclusive)
In addition to entering the threshold levels, tick the boxes next
to "Readings above" or "Readings below" to enable the alert
conditions for only the upper or only the lower threshold
settings, or for both. If both boxes are ticked, the behaviour is
the same as for the standard alert settings. The charts will
display alert readings according to the settings made in the
Charts (on page 90) section of the 'Options' Screen.
After X consecutive alert
readings (W minutes)
Readings above
Readings below
Tick the boxes next to "Readings above" or "Readings below" to
establish whether only readings above and including the upper
threshold or only readings below and including the lower
threshold should count towards the consecutive alert readings. If
both boxes are ticked, the behaviour is the same as for the
standard alert settings. If the Alert Indicator is disabled for either
the upper or the lower alert, the respective box for the
consecutive alert is grayed out and cannot be ticked.
After X alert readings have
occurred (W minutes)
Readings above
Readings below
Tick the boxes next to "Readings above" or "Readings below" to
establish whether only readings above and including the upper
threshold or only readings below and including the lower
threshold should count towards the total number of alert
readings. If both boxes are ticked, the behaviour is the same as
for the standard alert settings. If the Alert Indicator is disabled
for either the upper or the lower alert, the respective box for the
alert total is grayed out and cannot be ticked.
The advanced settings are displayed automatically without the need to press
if an advanced configuration is already found in the logger to be configured. If you do not
wish to use the advanced settings, press
, and the standard Alert Processing
(on page 34) configuration screen is displayed.
Example:
You wish to configure a LogTag so an alert is displayed if 5 continuous readings are
out-of-range or 7 total readings are out-of-range, but only if the temperature goes to
20ºC and above. You wish to see in LogTag Analyzer if the product has gone to 0ºC and
below, but you do not wish to trigger an alert for this condition. Alert parameters should
be set as follows: